19
Exposure Rate Calibraon
An exposure rate calibraon correlates the number of counts in a given me to the exposure rate present in
that eld. This is in many ways similar to eciency but we refer to it in this case as gamma sensivity. Although
it is possible to express this value in several dierent units, the Inspector EXP uses a formula of CPM/mR/hr
(counts per minute per milliroentgen per hour).
The instrument is placed in gamma radiaon elds of dierent (but known) intensies that span the working
range of the instrument, and the sengs are adjusted (if needed) such that all measurements are within ±10%
of the true value. The relevant sengs include both the sensivity and the dead me.
The gamma eld is usually generated by a relavely strong Cs-137 source. Exposure rates are calculated to the
center of the detector’s volume.
Requirements
You will need a PC running Observer USB and a Gamma calibraon range
Procedure
1. Power on the instrument.
2. Connect the instrument to the PC running Observer USB using the USB cable. An applicaon window for
the aached instrument should open within a few seconds.
3. Display the calibraon panel by clicking View|Cal Panel.
4. Record the displayed values for Dead Time and Sensivity.
5. Begin checking the instrument’s Sensivity. Place the instrument in a relavely low eld, approximately 5
mR/hr, and allow it 30 seconds or so to stabilize.
6. If the indicated response is o by more than 3% at this level, adjust the Sensivity to a lower value if the
instrument is reading too low, or to a higher value if the instrument is reading too high. Aer changing
the value on the Calibraon Panel, click the Update Sengs buon. Aer a second or two the instrument
will beep to indicate that the new value has been received.
7. Repeat Step 5 as needed to bring the instrument indicaon as close as possible to the true value.
8. Next, expose the instrument to a eld near the top of its range of 100 mR/hr. Allow the reading to
stabilize for at least 30 seconds. In the preferences tab, you can adjust the me in the gird to show a
selected average count.
9. If the indicated response is o by more than 10%, adjust the Dead Time to a lower value if the instrument
reading is too low, or to a higher value if the instrument reading is too high. Aer changing the value on
the Calibraon Panel, click the Update Sengs buon. Aer a second or two the instrument will beep
twice to indicate that the new value has been received.
10. Repeat step 9 as needed to bring the instrument’s indicaon within ±10% of the true value.
11. Once the values for Sensivity and Dead Time have been veried to be correct, expose the instrument
to a series of values spanning the range of the instrument. All indicaons should be within ±10% of the
true value.
Commentaires sur ces manuels