S.E. International Inspector EXP Spécifications Page 19

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 28
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 18
19
Exposure Rate Calibraon
An exposure rate calibraon correlates the number of counts in a given me to the exposure rate present in
that eld. This is in many ways similar to eciency but we refer to it in this case as gamma sensivity. Although
it is possible to express this value in several dierent units, the Inspector EXP uses a formula of CPM/mR/hr
(counts per minute per milliroentgen per hour).
The instrument is placed in gamma radiaon elds of dierent (but known) intensies that span the working
range of the instrument, and the sengs are adjusted (if needed) such that all measurements are within ±10%
of the true value. The relevant sengs include both the sensivity and the dead me.
The gamma eld is usually generated by a relavely strong Cs-137 source. Exposure rates are calculated to the
center of the detectors volume.
Requirements
You will need a PC running Observer USB and a Gamma calibraon range
Procedure
1. Power on the instrument.
2. Connect the instrument to the PC running Observer USB using the USB cable. An applicaon window for
the aached instrument should open within a few seconds.
3. Display the calibraon panel by clicking View|Cal Panel.
4. Record the displayed values for Dead Time and Sensivity.
5. Begin checking the instrument’s Sensivity. Place the instrument in a relavely low eld, approximately 5
mR/hr, and allow it 30 seconds or so to stabilize.
6. If the indicated response is o by more than 3% at this level, adjust the Sensivity to a lower value if the
instrument is reading too low, or to a higher value if the instrument is reading too high. Aer changing
the value on the Calibraon Panel, click the Update Sengs buon. Aer a second or two the instrument
will beep to indicate that the new value has been received.
7. Repeat Step 5 as needed to bring the instrument indicaon as close as possible to the true value.
8. Next, expose the instrument to a eld near the top of its range of 100 mR/hr. Allow the reading to
stabilize for at least 30 seconds. In the preferences tab, you can adjust the me in the gird to show a
selected average count.
9. If the indicated response is o by more than 10%, adjust the Dead Time to a lower value if the instrument
reading is too low, or to a higher value if the instrument reading is too high. Aer changing the value on
the Calibraon Panel, click the Update Sengs buon. Aer a second or two the instrument will beep
twice to indicate that the new value has been received.
10. Repeat step 9 as needed to bring the instruments indicaon within ±10% of the true value.
11. Once the values for Sensivity and Dead Time have been veried to be correct, expose the instrument
to a series of values spanning the range of the instrument. All indicaons should be within ±10% of the
true value.
Vue de la page 18
1 2 ... 14 15 16 17 18 19 20 21 22 23 24 ... 27 28

Commentaires sur ces manuels

Pas de commentaire